USER'S MANUAL
9. TIPS FOR OPERATORS AND SERVICE TECHNICIANS
Any repairs exceeding usual routine maintenance must
be carried out by suitably qualified personnel.
WARNING
Infection due to contaminated unit
Contaminated units can cause infections.
» Clean and disinfect the iM3 Bottle System before working on the unit.
» Wear suitable personal protective equipment when working
(e.g. impermeable gloves, protective goggles and mouth and nose protection).
ERROR
Low water flow rate
No water flow
Fluids leakage from the
bottle system
Air blows off audibly
10. MANUFACTURER OF BOTTLES
Manufactured and serviced by:
iM3 Dental Limited
Unit 29, Duleek Business Park,
Duleek, Co. Meath,
A92 N72W, Ireland.
Phone: +353 (0) 1 691 1277
Email: sales@im3dental.com |
support@im3dental.com
Website: www.im3vet.com
EVOLUTION VETERINARY DENTAL UNITS
POSSIBLE CAUSE
Fine filter blocked in the riser pipe
Blockage in the riser pipe
Blockage in the PUR hose
Kink in the PUR hose
Blockage in the riser pipe
Blockage in the PUR hose
No pressure
PUR hose is not attached correctly
Bottle not completely screwed in
Safety valve blows off
iM3 Inc.
12414 NE 95th St., Vancouver, WA,
98682, USA
Phone: +1 800 664 6348
Fax: +1 360 254 2940
Technical Support: +1 800 664 6348
Email: info@im3usa.com
Website: www.im3vet.com
SOLUTION
Replace the fine filter.
Remove and clean riser pipe.
Remove and clean the PUR hose, replace if
necessary.
Reposition the PUR hose.
Remove and replace riser pipe.
Remove and replace the PUR hose.
Switch on the compressed air supply.
Screw in the bottle as far as it will go, to open
the stop valve.
Reconnect the PUR hose with a hose safety
device.
Turn bottle to screw in completely.
Pull the ring of the safety valve to vent.