NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
►
Clean the product with a damp cloth only (fresh water).
INFORMATION
Movement noises of the joint unit and other joints (medial support, ankle joint)
When using orthotic joints, servomotor, hydraulic, pneumatic or brake load dependent control
functions can cause movement noises. This kind of noise is normal and unavoidable. It generally
does not indicate any problems. If movement noise increases noticeably during the lifecycle of
the orthosis, the orthosis should be inspected by an O&P professional immediately.
4.4 Information on the Power Supply/Battery Charging
CAUTION
Charging the product with damaged power supply unit/charger/charger cable
Falling due to unexpected behaviour of the product caused by insufficient charging.
►
Check the power supply unit, charger and charger cable for damage before use.
Replace any damaged power supply unit, charger or charger cable.
►
CAUTION
Charging while wearing the product
>
Falling due to walking and getting caught on the connected battery charger.
>
Falling due to unexpected product behaviour because of changed damping behaviour.
►
Only charge the product in a sitting position while wearing it.
NOTICE
Use of incorrect power supply unit/battery charger
Damage to product due to incorrect voltage, current or polarity.
Use only power supply units/battery chargers approved for this product by Ottobock (see
►
instructions for use and catalogues).
4.5 Battery charger information
NOTICE
Penetration of dirt and humidity into the product
Lack of proper charging functionality due to malfunction.
Ensure that neither solid particles nor liquids can penetrate into the product.
►
NOTICE
Mechanical stress on the power supply/battery charger
Lack of proper charging functionality due to malfunction.
►
Do not subject the power supply/battery charger to mechanical vibrations or impacts.
►
Check the power supply/battery charger for visible damage before each use.