Directives And Manufacturer's Declaration; Electromagnetic Environment - Ottobock C-Brace 17KO1 Serie Instrucciones De Uso

Ocultar thumbs Ver también para C-Brace 17KO1 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 207
LED on
LED on bat­
power
tery charger
supply

15.3 Directives and manufacturer's declaration

15.3.1 Electromagnetic environment

This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see
Page 63).
The following tables outline the test levels of the tests to be carried out. The higher value applies if
there is a difference between the operating environments listed in the tables below.
Table 1 – Electromagnetic emissions for all devices and systems
Interference measurements
HF emissions according to
CISPR 11
HF emissions according to
CISPR 11
Harmonics according to
IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
98
Error
Battery is fully charged (or con­
nection with product is interrup­
ted).
Compliance
Group 1
Class B
N/A
N/A
Resolution
Take note of the confirmation sig­
nal for differentiation.
When the battery charger is con­
nected, a self-test is conducted
and confirmed by beeps/vibration
signals.
When these signals are emitted,
the battery is fully charged.
If no signals are emitted, the con­
nection to the product is interrup­
ted.
If the connection to the product is
interrupted, an authorised
Ottobock Service Centre must
inspect the product, battery char­
ger and power supply.
Electromagnetic environ­
ment directive
The product uses HF energy
exclusively for its internal func­
tioning. Its HF emissions are
therefore very low, and inter­
ference with neighbouring
electronic devices is unlikely.
The product is intended for
use in other than residential
facilities and in facilities con­
nected directly to a public low-
voltage network that also sup­
plies residential buildings.