Section X - Specifications And Symbols; Specifications - Coviden Kangaroo ePump Manual De Instrucciones

Bomba para irrigación y alimentación enteral con abrazadera para palo de gotero, programable
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 197
Section X — Specifications and Symbols

Specifications

TUV Listing
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump Sets
Kangaroo™ ePump™ Feed-Only Set or Feed & Flush Set (with MISTIC Connector)
Feeding Formula Delivery Rate
1-400 mL/hr in 1 mL increments
Feeding Formula VTBD
1-3000 mL in 1 mL increments
Bolus Volume
1-3000 mL in 1 mL increments
Number of Boluses
1-99
Bolus Interval
1-24 hours in 1-hour increments
Flushing Solution Dose Range
10-500 mL in 1 mL increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
±10% or 0.5 mL/hr, whichever is larger, with bag at 46 cm (18") above pump, at room temperature 22° C ± 2° C (72° F ± 3° F), using
water and a new pump set with less than the recommended 24 hours of maximum usage.
Single Fault Condition Effect on Accuracy - In the event of a single fault short circuit in the pump electronics, a maximum bolus of
1.67 mL of extra fluid delivery may occur during feeding and 8.19 ml during flushing or priming.
Feeding Set Occlusion Pressure
15 psi (103 kPa) Nominal
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Medical Electrical Equipment
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
(1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with
ES60601-1:2012, UL60601-1, IEC 60601-1:2012.
(2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance
with CAN/CSA C22.2 No. 60601-1:08.
31