The Kangaroo™ ePump™ enteral feed and flush pump with pole clamp are intended for use in the electromagnetic environment specified
below. The user of the Kangaroo™ ePump™ enteral feed and flush pump with pole clamp should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
(EN 61000-4-2 per
EN 60601-1-2: 2015)
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
(EN 61000-4-8 per EN
60601-1-2: 2015)
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601 Test Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV differential mode
Nominal Mains Voltage
(VNOM) 100 VAC and 240 VAC
@ 50 Hz
Voltage Dips
100% of VNOM for 10 mSec (0.5
Line Cycles) at 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
100% of VNOM for 20 mSec
(1 Line Cycle) at 0°
70% of VNOM for 500 mSec (25
Line Cycles) at 0°
Interruptions
100% of VNOMº for 5000 mSec
(250 Line Cycles)
30 A/m
Compliance Level
± 8 kV contact
± 15 kV air
±2 kV for power supply lines
± 1 kV differential mode
No degradation of performance
or loss of function.
30 A/m
39
Electromagnetic Environment -Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Kangaroo™
ePump™ enteral feed and flush pump with
pole clamp require continued operation
during power mains interruptions, it is
recommended that the device be powered
from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.