Section Xv - Electromagnetic Conformity Declaration - Coviden Kangaroo ePump Manual De Instrucciones

Bomba para irrigación y alimentación enteral con abrazadera para palo de gotero, programable
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 197
Section XV – Electromagnetic Conformity Declaration
The Kangaroo™ ePump™ enteral feed and flush pump with pole clamp has been built and tested according to UL 60601-1 (2nd
Edition), ES60601-1:2012, CAN/CSA C22.2 No. 60601-1-08, and EN60601-1-2 Standards.
The Kangaroo™ ePump™ enteral feed and flush pump with pole clamp is intended for use in the electromagnetic environment
specified below. The user of the pump should assure that it is used in such an environment.
The Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp is intended for use in the electromagnetic environment specified
below. The user of the Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp should assure that it is used in such an environment.
Emissions Test
RF emissions
(CISPR 11)
RF emissions
(CISPR 11)
Radiated Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-3: 2006 + A1: 2007 +
A2: 2010)
Conducted Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-6:2013)
Power Frequency Magnetic Field Immunity
(EN60601-1-2 / IEC 61000-4-8:2009)
Voltage dips and sags Immunity
(EN60601-1-2 / IEC 61000-4-11:2004)
Electrical Fast Transient / Bursts Immunity
(EN60601-1-2 / IEC 61000-4-4:2012)
Electrostatic Discharge Immunity
(EN60601-1-2 / IEC 61000-4-2:2008)
Surge Immunity
(EN60601-1-2 / IEC 61000-4-5:2005 +Cor 1: 2009)
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Compliance
Group 1
Class B
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Electromagnetic Environment - Guidance
The Kangaroo™ ePump™ enteral feed and flush pump uses RF
energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby
electronic equipment.
The Kangaroo™ ePump™ enteral feed and flush pump is suitable for
use in all establishments, including domestic establishments and
those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.