Section VIII — Battery Replacement
See Figure 24 and do the following to replace the battery pack:
• Power down the unit.
• Loosen the battery door screw, on back of pump.
• Open battery door and disconnect wire harness. A small straight-bladed screw driver may be needed to disconnect the
connector tab.
• Slide battery pack out of compartment.
• Slide a new battery pack into compartment with wire harness
extending out.
• Align connector with red wire at upper right.
• Push connector in until properly latched.
• Caution: Ensure the battery wires are routed through the slot
between the connector and the battery to prevent the wires
from becoming pinched.
• Re-tighten the battery door screw; if there is any resistance
while securing the door, recheck the wires to verify they are
properly routed through the wire slot.
Notes:
• Replacement battery packs are available from Covidien (See
Section XIII - Service Part Numbers).
• The battery is packed partially charged. Charge battery 6 hours prior to use off A/C power.
• To assure the safe and reliable operation of this pumping system, do not attempt to operate without an approved battery pack
installed in the pump.
• Dispose of used battery cells in accordance with local and facility guidelines.
• Keep away from children.
Warning: Use of a battery pack not supplied by Covidien can be dangerous and voids all warranties and
performance specifications.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully
recharge the battery pack.
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Figure 24. Installation of Battery Pack.
24