The Kangaroo™ ePump™ enteral feed and flush pump with pole clamp are intended for use in the electromagnetic environment specified below.
The customer or the user of the Kangaroo™ ePump™ enteral feed and flush pump with pole clamp should assure that it is used in such
an environment
Immunity Test
Conducted RF
6 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
Band
(EN 61000-4-3 per
380-390 MHz
EN 60601-1-2: 2015)
430-470 MHz
704-787 MHz
800-900 MHz
1700-1990 MHz
2400-2570 MHz
5100-5800 MHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Kangaroo™ ePump™ enteral feed and flush pump is used exceeds the applicable RF compliance level above, the Kangaroo™ ePump™ enteral feed
and flush pump should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Kangaroo™
ePump™ enteral feed and flush pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC 60601 Test Level
Frequency Modulation
385 MHz
PM, 18 Hz, 50% 27 V/m
450 MHz
PM, 18 Hz, 50% 28 V/m
710 MHz
PM, 217 Hz, 50% 9 V/m
745 MHz
780 MHz
810 MHz
PM, 18 Hz, 50% 28 V/m
870 MHz
930 MHz
1720 MHz
PM, 217 Hz, 50% 28 V/m
1845 MHz
1970 MHz
2450 MHz
PM, 217 Hz, 50% 28 V/m
5240 MHz
PM, 217 Hz, 50% 9 V/m
5500 MHz
5785 MHz
Compliance
Level
6 Vrms
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
40
Electromagnetic Environment -Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Kangaroo™ ePump™ enteral
feed and flush pump with pole clamp,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
Not applicable
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency
range. b Interference may occur in the
vicinity of equipment marked with the
following symbol: