Section V - Instructions For Use; Quick Start - Coviden Kangaroo ePump Manual De Instrucciones

Bomba para irrigación y alimentación enteral con abrazadera para palo de gotero, programable
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 197
Section V — Instructions For Use

Quick Start

POWER on lower, right-hand corner of control panel.
1. Press
2. Open blue door enclosing pump set loading area.
3. To load pump set , refer to Figure 4:
• Grasp finger tab on valve and insert into left pocket (1).
• Grasp black ring retainer and gently stretch tubing around rotor (2). Pull up on retainer and insert into right pocket (3).
• Press down on finger tab to ensure that valve is fully seated. Finger tab should align with locating rib.
• Close blue door.
(Note: display status line should read SET LOADED).
4. To automatically prime the pump, press
priming will prime both lines.
5. To directly control pump priming, press
& flush set is used, remember to first press
press
HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the distal end connector.
4
6. Use buttons
on left side of LCD to set the feeding parameters. Set flushing parameters if a feed & flush set is loaded. The
4
continuous or intermittent feed mode selection is made under the MORE options menu.
7. Press
RUN when ready to start. The screen will display RUNNING.
4
HOLD or press and hold the
8. To stop, press
4
Kangaroo™ ePump™ Enteral Feed and Flush Pump with Pole Clamp
PRIME PUMP and
4
PRIME PUMP and then press and hold the HOLD TO PRIME button(s)
4
HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then
4
POWER button.
Finger
Tab
1
Valve
2
Figure 4. Installation of pump set.
AUTO PRIME. For Kangaroo™ ePump™ feed & flush sets, auto
4
Black Ring
Retainer
(MISTIC)
3
Rotor
8
. If a feed
4
Table of Contents

Publicidad

Tabla de contenido
loading

Tabla de contenido