Error #
Related Sub-system
0
Internal timer stopped.
1
Flash Chip missing OR version incompatible.
2
Too many graphic screens in memory.
4
Never got a ADC ISR for motor current.
5
Never got a ADC ISR for battery level.
6
Never got a ADC ISR for battery temperature.
7
Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-Empty checks.
8
Never got a ADC ISR for Ultrasonic during Downstream Occlusion.
9
RS232 Communications Transmission Error.
10
Flash Write Error.
11
Flash Read Error.
12
General Stopcock Time Out.
13
Time Out for Buzzer Test
16
Flash Screen Read Error.
19
Flash Languages Read Error.
20
Could not find Stop Cock Flush Position during prime routine.
21
Could not find Stop Cock Feed Position during prime routine.
22
Motor Error during Tubing AutoLoad.
23
Motor Failsafe Time Out Error. Possible damaged gearbox encoder or flex cable.
25
Display Lockup Time Out Error.
33
Motor Control Failure.
Turn the power off and back on to try to clear the error. If the error cannot be resolved in a timely manner, press
operation of the pump and put a different pump into service.