Directions for Cleaning Pump Housing
• Refer to General Cleaning Directions before starting.
• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as possible.
• Avoid excess moisture near pole clamp area.
• Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.
Directions for Cleaning Pump A/C Power Adapter
• Refer to General Cleaning Directions before starting.
• Unless soiling is observed, the A/C power cord should not be cleaned.
• If cleaning of the power adapter is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with a cloth
dampened with isopropyl alcohol.
Caution: Avoid exposing A/C power adapter to excess moisture, as this can lead to an electrical shock or fire hazard.
Directions for Cleaning Rotor Assembly
• Refer to General Cleaning Directions before starting.
• Open the blue door enclosing the pump set loading area.
• Loosen rotor set screw with 5/64" (2 mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid getting any
moisture in the rotor shaft opening.
• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
• Be sure all parts of rotor are completely dry before putting it back onto shaft.
• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
• Push rotor into place and tighten set screw (Do not over tighten).

Preventative Maintenance

This pump must be periodically tested to assure proper functioning and safety. The recommended service interval is every 2 years.
Performance testing may be done at the user's Biomedical Engineering Department, an outside service, or by Covidien Factory
Service. To arrange for Covidien Factory Service in the US, call 1-800-962-9888. In Canada, call 1-877-664-8926. Outside of the US
and Canada, please contact your local Customer Service for more information.
If a pump malfunctions, please contact your Covidien Representative or call Customer Service for instruction.