Guidance and manufacturer's declaration – Electromagnetic immunity
hyperion MRT / hyperion is intended for use in the electromagnetic environment specified below. The customer or
the user of the hyperion MRT / hyperion should assure that it is used in such an environment.
Immunity Test
Radiated RF
EN 61000-4-3
Conducted RF
EN 61000-4-6
Recommended separation distances between portable and mobile RF communications equipment and
hyperion MRT / hyperion is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the hyperion MRT / hyperion can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitter) and the hyperion MRT / hyperion as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output power of the
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at the maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
(1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
EN
IEC 60601-1-2
Test level
3 V/m
80 MHz to 2.5 GHz
3 V
150 kHz to 80 MHz
the hyperion MRT / hyperion.
Separation distance according to frequency of transmitter
150KHz to 80MHz
d = 1.2 x √P
OPERATOR'S MANUAL
Compliance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of the hyperion MRT / hyperion,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance.
d = 1.2 x √P
3 V/m
d = 2.3 x √P
d = 1.2 x √P
3 V
Where P is the maximum output rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended
meters (m).
Field strengths for fixed RF transmitter, as
determined by an electromagnetic site
survey, should be less then the compliance
level in each frequency range interference
may occur in the vicinity of equipment
marked with the following symbol:
80KHz to 800MHz
d = 1.2 x √P
0.12
0.38
1.2
3.8
12
Electromagnetic Environment
80 MHz to 800MHz
800 MHz to 2.5GHz
separation
(m)
800KHz to 2.5MHz
d = 2.3 x √P
0.12
0.38
1.2
3.8
12
distance
in
0.23
0.73
2.3
7.3
23
13