Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Only CAIMAN instruments can be used in conjunction with the Lektrafuse
HF generator GN200, see Accessories/Spare parts.
Connecting the power supply
Risk of death from electric shock!
Only ever connect the product to power mains
DANGER
with equipment grounding conductor.
Set up the device in such a way that switching
it off using the power OFF/ON switch is possible
and straightforward.
Set up the device in such a way that disconnect-
ing the power cord is straightforwardly possible.
The voltage of the mains power supply must correspond to the voltage
indicated on the type plate of the device.
The device is fitted with a universal power supply unit, which means that
it automatically adapts to mains voltages ranging from 100 V to 240 V
without having to switch between voltage ranges.
Ensure that the device is switched off. Also switch off with the power
OFF/ON switch 12.
Plug in the power cord in mains power input socket 13 in the rear panel
of the HF surgical device.
Plug in the mains plug at the building mains power socket.
Switching on the device
Switch on the device with the power OFF/ON switch 12.
Indicator lamp power ON 9 is illuminated. The device performs a self-
test, see Continuous test cycle during operation:
If no instrument is connected, the message "Attach Instrument"
appears in the display 5 after the self-test. If the self-test fails, an error
message is shown in the display 5.
Fig. 4
Connecting the foot control
Connecting the foot switch GN201 is optional.
Both the foot switch plug and connection socket (foot switch) 3 have a
dot marking. To connect the plug to the connection socket correctly, these
markings must be aligned.
Fig. 5
Align the foot switch plug and insert it into the foot switch connection
socket 3 until it engages.
When the device and the foot switch have been successfully connected,
the message "Footswitch attached" appears for 3 seconds in the display
5.
Fig. 6
Note
Do not activate the foot switch continually during HF discharge. HF dis-
charge is started and stopped by pressing once on the foot switch and then
releasing.
Remove the foot switch
Turn the locking ring counter-clockwise and withdraw the connecting
plug at the same time
The message "Footswitch removed" appears in the display 5 for 3 sec-
onds.
Fig. 7
en
7