6.4 Features of alarm signals
Alarm
Priority
Visual Signal
Alarm indicator is in Red.
Cycle: 500ms.
High
The alarm cause is
displayed.
Alarm indicator is in
Yellow.
Cycle: 2s.
Middle
The alarm cause is
displayed.
Alarm indicator is in
Yellow.
The indicator lights
Low
continuously.
The alarm cause is
displayed.
Note: There is 55-82dB(A) sound pressure caused when 1 meter far from the machine.
Information:
If Battery Empty alarm or No Battery alarm is activated, the battery symbol will
be also flashing on the display.
6.5 Alarm silence
Press
key to silence an auditory alarm.
Information:
The following alarms cannot be silenced: Key Holding During Infusion / Pause
Out of Flow Range, Out of Dose Range, Battery Low, Battery Empty and No
Battery.
The visual signals including both indicator light and displayed message are still
visible when an auditory alarm is silenced.
The silenced auditory alarm resumes in 110 seconds or if
again.
Battery Empty alarm is prior to Rate Abnormal alarm.
Battery Empty alarm is prior to Key Holding During Infusion alarm.
Auditory Signal
5 tones, repeat.
5 tones, repeat.
2 tones, Cycle: 20s
Alarm indicator
flashing
frequency
2.0± 0.6HZ
0.6± 0.6HZ
Always
Duty cycle
20%`60%
20%`60%
100%
,
key is pressed
102