won't topple over when tilted by 20 degrees.
3.3 Stocking the pump
Up to 8 pumps can be stacked up on one pole as stackable
syringe pumps.
5.
Use the pole clamp to mount the first pump on the pole
securely.
6. Place the second pump on first one by the stacking
clamping groove as chart A.
7. Repeat the step1 to 2 to stack more pumps as required
8. Screw the pole clamp to mount on the pole securely as Chart
B.
3.4 Loading the syringe
3.4.1 Syringe recommendation
Warning:
The syringes loaded on the device should comply with the standard ISO
7886-1,and we suggest use of Luer syringes.
Use of syringes that do not meet the above requirements may cause incorrect
flow rate or other damages to the device and the patient.
3.4.2 Loading procedure
Warning:
Check the syringe, and it must be calibrated before use as below situation
details in Chapter 4.3 "Syringe calibration"
-
-
-
The brand of syringe used is not the own brand supplied by the pump.
The brand & size of syringe is new.
Significant residual liquid is found left in the syringe when infusion is
):
(
Under the condition of user defined
Chart A
Chart B
see
(
)
76