Invalid time is displayed
and the time cannot be
adjusted.
History records cannot
be exported.
The displayed syringe
size is incorrect.
The syringe calibration
cannot be performed.
Displayed pressure level
is obviously disagrees
with reality.
7.2 Error messages
Information:
Error messages including the fault and alarm information caused in power-on
and run-time self-test, these all messages displayed in the form of text and ICONS
on the screen.
Message displayed
EEPROM
Battery loose
Syringe calibration
error
Pressure sensor
error
Main board is defective.
Button cell is exhausted.
Main board is defective.
RS232 circuit on main board is
defective.
RS232
connection
disconnected.
Syringe size detection board is badly
connected.
Syringe size detection board is not
installed properly or is defective.
Main board is defective.
Potentiometer connected to J6 on main
board is disconnected or defective.
Main board is defective.
Force sensor connection wires are
disconnected.
Force sensor is defective.
Main board is defective.
Description
EEPROM chip or mainboard
failure
No battery in machine
Battery failure
Battery connection wires are
shorted.
The
EEPROM
calibration syringes dates are
failure
Incorrect voltage presents at
point TP7 on main board
when syringe is not installed
or main board is defective
Replace main board.
Replace button cell.
Replace main board.
Replace main board.
wires
are
Reconnect RS232 port to main
board.
Reconnect
related wires or connectors.
Reinstall or replace syringe
size detection board.
Replace main board.
Reconnect
potentiometer.
Replace main board.
Reconnect force sensor.
Replace force sensor.
Replace main board.
Recommended actions
Replace
mainboard
Install battery on mainboard
Replace battery
storage
Calibration the syringe again
If main board or pressure sensor is
replaced, adjust potentiometer PJ1
and make the voltage at TP7 on
main board within range of
or
replace
the
or
replace
EEPROM
chip
or
106