Out of Dose Range
Information:
The device stops infusion and triggers audio and video alarm signals when a
high or medium priority alarm occurs.
The device goes to KVO mode automatically when Infusion Complete alarm is
activated.
6.2 Alarm conditions and corrective actions
Alarm
Syringe Holder Loose
Syringe Plunger
Disengagement
Syringe Flange
Insertion Error
Syringe Not Calibrated
Occlusion
Syringe Empty
Rate Abnormal
Battery Empty
No Battery
Key Holding During
Infusion/ Pause
L
Yellow
Alarm Condition
Syringe holder is turned and not
holding the syringe any more during
infusion.
The syringe release lever of the pusher
is pressed during infusion.
Syringe flange is not detected during
infusion.
Syringe is not calibrated before it is
loaded.
Pressure inside the syringe and
extension line reaches a certain level
(H, M or L) during infusion.
The solution in the syringe has been
infused completely and the syringe
becomes empty.
The detected syringe plunger
displacement differs from the expected
one.
The remaining battery life is less than
3 minutes.
No battery or the battery is dead
A key is stuck or was held for over 5
seconds during infusion or pause.
Suggested action
Check the syringe installation
and ensure the syringe holder is
in the right position and holds
the syringe.
Check the syringe installation
and ensure the syringe is in
close contact with pusher and
syringe lever is not pressed
during infusion.
Check the syringe installation
and ensure the flange is placed
in the flange slot and the
syringe holder is in right
position.
Calibrate the syringe.
Check the syringe and
extension line for signs of
occlusion and take necessary
actions.
Replace syringe or end the
infusion.
Recalibrate the syringe.
Connect to the AC power
source.
Contact manufacturer.
Check and release the key
Y
100