Information:
 After infusion is stopped:
-
-
4.3.3 Rate mode
Information:
 In Rate Mode, the parameters of flow rate ,VTBI and occlusion alarming
pressure level can be set by the user.
 Press
key till the display shows "MODE: RATE".
 Press
or
when the parameter is selected, it will be highlight displayed.
 Use
or
Information:
 The occlusion alarming pressure level:
High: 120(±15) kPa.
Medium: 90(±15) kPa.
Low: 60(±15) kPa.
It should be set according to the clinical needs and under the guidance of
professional personage 
 Press
key to start infusion.
Information:
 To adjust flow rate during infusion, please refer to section "Change flow rate
setting during infusion" in this chapter.
is displayed when infusion is stopped.
User may adjust parameters and restart infusion again.
The infused volume will be not automatic clear zero, if need
.
key to select the parameter to set ( "Rate" "VTBI" 、 "Occ" ( H,M,L )) ,
to set the parameter value as required.
please press
83