8.1.6 KVO Rate
0.1 ~ 5.0mL/h or the last delivery rate whichever is smaller.
8.1.7 The occlusion alarming pressure level
High:
120(±15) kPa.
Medium:
90(±15) kPa.
Low:
60(±15) kPa.
8.1.8 The details about the Occlusion alarm
Information:
The maximum occlusion pressure in the syringe and extension line generated by
the device is 250kPa
When an occlusion alarm occurs, anti-bolus function is started automatically to
reduce the bolus generated by the occlusion
8.1.9 BOLUS volume released by Anti-bolus
Flow Rate
1mL/h
5mL/h
8.1.10 Parameters of Dose mode
Range
0.001 ~ 9.999, Increment 0.001
10.0 ~ 99.99, Increment 0.01
Dose
10.0 ~ 99.99, Increment 0.01
100.0 ~ 999.9, Increment 0.1
1000 ~ 9999,
ng/kg/min,ug/kg/min,mg/kg/min,g/kg/min,ug/kg/h,mg/kg/h,g/kg/h,
Dose unit
ug/min,mg/min,g/min,ug/h,mg/h,g/h,U/min,U/h,IU/h,U/kg/h,ml/h
Weight
0.1 ~ 300.0kg, Increment : 0.1kg
Drug mass
0.1 ~ 999.9mg, Increment : 0.1mg
Flow Rate
1ml/h
5ml/h
。
Increment 1
Occlusion level
High
Low
High
Low
Occlusion level
High
Low
High
Low
Alarm delay time
≤2.5h
≤2h
≤30min
≤30min
BOLUS volume
≤0.5mL
≤0.3mL
≤0.5mL
≤0.3mL
110