The start-up curve represents the change in flow rate over a certain period of time, and the
trumpet curve shows the evolution of the minimum and maximum variance of the extension line,
syringe and pump combination.
The following graphs are obtained from the tests performed with the system consisting of syringe
pump and Come Sure 50ml syringe. The fluid used is distilled water.
These tests were conducted in accordance with IEC60601-2-24 requirements.
The results may not be very close to those obtained in clinic because of the variations in extension
lines, syringes, physical properties of infused fluid, and the environment etc. The error in flow
accuracy caused by such variations may be larger than 5%
8.3.1 Start-up curves
Flow rate
: 5mL/h
Flow rate
: 1mL/h
Compliant
with
60601-1:2006,
60601-2-24:1998,
60601-1-2:2007/AC2010,
IEC
60601-1-8:2008,
IEC 60529:2001 standards.
Compliant
with
60601-1-2:2007/AC2010 and
EN/IEC 60601-2-24:1998
standards
IEC
IPX4
IEC
IEC
and
EN/IEC
Protection against splashing
liquid.
Protection
against
current:
Defibrillation-proof
type CF applied part.
Protection
against
shocks: class I
Functional earth
leakage
electric