Pressure
calibration error
KEY ERR
Rate abnomal
FLANGE
INSERTION
ERROR
PLUNGER
DISENGAGED
LOOSE
OCCLUSION
EMPTY
BATT EMPTY
Pressure
sensor
disconnected
Pressure sensor is defective
Pressure is not calibrated or
pressure calibration error.
Key holding during infusion /
pause alarm
A key is stuck or was held for
over 5 seconds.
Actual flow rate is
comply with the expected
rate.
Potentiometer
wires are shorted.
Motor running check the
sensor failure
Syringe
flange
board is defective
Syringe
flange
board is disconnected from
main board
Main board is defective
Syringe
plunger
board
or
plunger
detection board is defective
Plunger lock detection board
is disconnected from main
board
Main board is defective
Syringe loose detection board
is defective
Syringe loose detection board
is disconnected from main
board
Main board is defective
Syringe or extension line is
occluded
Pressure sensor is defective
Syringe is empty
Battery is exhausted
is
200±5mV when syringe is not
installed.
Reconnect pressure sensor
Replace pressure sensor
Perform pressure calibration
Check keyboard and release the
stuck key.
Replace the keyboard
not
Calibration syringes
Replace motor
Re-connection
connection
potentiometer
detection
Replace syringe flange detection
board
detection
Re-solder the wire connecting
syringe flange detection board
and main board
Replace main board
detection
Replace
lock
detection board
Replace plunger lock detection
board
Re-solder the wire connecting
plunger lock detection board and
main board
Replace main board
Replace syringe loose detection
board
Re-solder the wire connecting
syringe loose detection board and
main board
Replace main board
Check syringe or extension line
for sign of occlusion and take
necessary connective action
Replace pressure sensor
Change syringe or end the
infusion
Connect to AC power source
or
replace
syringe
plunger
107