6.6 Alarm clearing
 Press
key to clear an activated alarm.
Warning:
 After an alarm is cleared, the infusion is still stopped, and both auditory signal
and visual signal disappear. But the alarm cause still needs to be found and
corrective action needs to be taken.
Information:
 Press
Infusion / Pause
Empty and No Battery.
6.7 Characteristics of auditory alarm signals
 The high priority alarm signal sound pressure level >= the medium priority alarm signal sound
pressure level >= the low priority alarm signal sound pressure level.
 High priority auditory alarm signals of a particular set of alarm signals shall convey a higher
level of urgency than the low priority alarm signals and information signals of that alarm signal
set.
 Medium priority auditory alarm signals of a particular set of alarm signals shall convey a higher
level of urgency than the low priority alarm signals and information signals of that alarm signal
set.
6.8 Alarm delay
The mean alarm condition delay plus the mean alarm signal delay is greater than 5 seconds.
key, the following alarms cannot be cleared: Key Holding During
Out of Flow Range, Out of Dose Range, Battery Low, Battery
103