4.3.6 Dose mode
Press
key till the display shows "MODE: DOSE".
Press
key return to Rate mode interface.
Press
or
"SOL VOL", ) , when the parameter is selected, it will be highlight displayed..
 Use
or
Information:
 Dose range
 Available dose units include
U/kg/min, IU/kg/min, ng/kg/h, ug/kg/h, mg/kg/h, g/kg/h, U/kg/h, IU/kg/h,
ng/min, ug/min, mg/ min, g/min, U/min, IU/min, ng/h, ug/h, mg/ h, g/h, U/h,
IU/h , ml/h
 "Drug Mass" is the mass of the drug being dissolved in the solution.
 Drug Mass range
 "Weight" may not be available for setting for some dose units.
 Weight range
 The solution volume range
 The flow rate value changes with the above parameter settings.
Warning:
 Improper parameter settings may cause the set flow rate out of limit and trigger
the "Out of Flow Range" alarm.
When select
Information:
Select
and the parameter was set and effective.
Select
key to select the parameter to set ( "Dose", "Dose unit", "Weight", "Drug Mass",
to set the parameter value as required.
0.001
9999
。 
。 
0.1
999.9
0.1
300.0 kg
or
, press
and press
and press
ng/kg/min, ug/kg/min, mg/kg/min, g/kg/min,
in units
g,
。 
0.1
9999.9 ml
。 
will be disappeared.
, user interface will be changed.
, the display changes into Dose mode interface
, the display changes into rate mode interface.
mg, U, IU
。 