Press
/
selected, it will be highlight displayed.
After select "Check",
Information:
The history records including: start date&time, ending date&time, flow rate, infused
4.4.1.11 Brightness
Information:
Setting the brightness to adapt to the clinical use of the environment.
Press
key to enhance the display brightness.
key to weaken the display brightness.
Press
4.4.1.12 Software version check
To check the software version and hardware version of the pump.
4.5 Device reset
Insert the end of a needle or a straightened paper clip or
a toothpick into the reset hole near the AC power inlet.
Push it as far as it goes till the display turns black.
Information:
The device needs to be reset when the control of the device cannot be restored by
any other means, for example, the device cannot be switched off.
Warning:
Contact the manufacturer if the reset does not work and the device
malfunctioning persists.
key to select the parameter to set ( "Check", "Export" ) , when the parameter is
press
key to enter into history record interface.
volume, and stop position.
Use
or
key to check page by page.
97