 The total infused volume displayed is the current bolus volume infused.
Keep pressing
 Release
bolus function.
Information:
Bolus function continues as long as
provided no alarm comes in between or the target bolus volume limit is not reached.
 If the infused bolus volume reaches the target bolus volume, the device stops
bolus function and resumes the prior infusion, and the audio alarm tone also
sounds once.
 The bolus rate resumes its default value of 100ml/h when the device is restarted.
 In Automatic Bolus mode 。 
Press and release
blinking.
Press
key again immediately to start bolus function
Information:
 After bolus function is started:
rate resumes the prior infusion, till the infused volume reaches target volume,
the current Bolus infusion cycle is finished and to start the next cycle.
down, provided no alarm comes in between or the target bolus volume limit is
not reached.
4.3.9
Purge
Information:
-
key as long as you want the bolus to last.
key to stop the bolus function and resumes the infusion prior to starting the
key , BOLUS function is activated and"BOLUS" is displayed and
 Preset bolus flow rate, instead of normal flow rate, is displayed.
The total infused volume displayed is the current bolus volume infused.
If the infused bolus volume reaches the target bolus volume, the flow
 Bolus function continues as long as
Purge function is used to drive air out of the extension line quickly
key is pressed and held down,
key is pressed and held
89