Warning:
 Pause the infusion before start the Purge function.
 Ensure that the extension line is not connected to the patient during purging.
 When occlusion occurs in the extension line, do NOT start purge function,
instead perform purge manually.
 Pause the infusion and
 Press and release
 Press and hold
Information:
After purge is started:
 Purge continues at factory default flow rate as specified in Chapter 8.
 Release
4.3.10 KVO (
Keep Vein Open
When the infused volume reaches the target volume during infusion, the normal flow rate
automatically switches to KVO rate.
Information:
KVO rates: from 0.1 to 5.0ml/h (Set by user, see chapter 4.4.13) or the last
delivery rate, whichever is smaller.
 After KVO is started:
-
-
KVO continues until
4.3.11 Change flow rate setting during infusion
Information:
 This function is available only when infusion is in rate mode.
Start infusion in rate mode, press
is displayed. 
key, and
key immediately to start purge.
is displayed and blinking. The triangles of the operation indicator
become green in turn repeatedly.
key to stop the purge function and the device returns to be paused.
)
The audio alarm is activated.
The KVO indicator is blinking.
key is pressed or any higher priority alarm interrupts.
is displayed.
key once to unlock other keys
, the flow rate displayed 
90