8 Technical Information
8.1 Performance
8.1.1 Infusion flow rate
Range: 50/60mL syringe : 0.1 ~ 2000.0 mL/h
30mL syringe : 0.1 ~ 1200.0 mL/h
20mL syringe : 0.1 ~ 800.0 mL/h
10mL syringe : 0.1 ~ 400.0 mL/h
5mL syringe : 0.1 ~ 200.0 mL/h
Increments : 1mL/h or 0.1mL/h
Accuracy : ± 3% ( After corrected calibration, Mechanical Accuracy : ± 1% , Flow Accuracy :
Explain : When flow rate<5mL/h, the accuracy will reduce. It is recommended to use a smaller
size of syringe, and users shall carefully monitor the actual flow rate.
Warning
The accuracy will be influenced by such factors as the configuration of extend
tube, liquid viscosity coefficient and operational environment.
8.1.2 Volume to be infused
Range: 0.0 ~ 9999.9mL
Increment:
0.1mL
8.1.3 Infusion Time
Range :
00h01min ~ 99h59min
Increment : 1min
8.1.4 Bolus or Purge rates
Syringe
50 /60mL
30 mL
20 mL
10 mL
5 mL
8.1.5 BOLUS volume
Range :
1.0 ~ 50.0mL
Increment : 0.1mL
± 2% )
:
(Increment:0.1 mL/h)
0.1 to 2000.0 mL/h
0.1 to 1200.0 mL/h
0.1 to 800.0 mL/h
0.1 to 400.0 mL/h
0.1 to 200.0 mL/h
BOLUS Rate
Purge Rate
2000.0 mL/h
1200.0 mL/h
800.0 mL/h
400.0 mL/h
200.0 mL/h
109