TO ADJUST THE ACTIVE THERAPY PARAMETERS
1 .
Select the "Vibration" button .
2 .
Adjust the therapy parameter as described in
step 2 of section "To Start Vibration Therapy"
on
page
1-21 .
Note:
• The adjusted parameters take effect as soon as
they are changed .
• To change lung selection, Vibration must be
stopped .
• The vibration therapy session continues until
the selected duration time is reached or until the
session is stopped or paused .
• When restarting vibration therapy after the session
was paused or stopped for more than 30 seconds,
the screen will display two options: "Continue
Previous Therapy Session" or "start New Therapy
Session" .
• All therapy parameters are stored until changed or
reset .
• To reset the parameter to the default values, refer to
"Advanced Settings" on
• The Percussion therapy button is inactive while
Vibration is active .
TO STOP VIBRATION THERAPY
1 .
Select the "Stop" button and the
Vibration therapy is stopped .