TO START MAX INFLATE

The Max Inflate mode inflates the air bladders to the
maximum air pressure, creating a firm surface for patient
repositioning, transfer, or CPR . Max Inflate should not be
used as a firmness setting and is only intended for a
short duration of up to 30 minutes .
1 .
Select the "Max Inflate" button to start .
Note:
• The Rotation, Percussion, Vibration, Firmness and
Turn Assist buttons are all inactive when Max Inflate
is active .
• If Rotation, Percussion or Vibration are active when
Max Inflate is selected, the therapy will pause and
can be restarted by pressing "Start" after Max
Inflate finishes or is stopped .
• After 30 minutes of continuous Max Inflate operation,
the message "Maximum Inflate Has Reached Time
Limit" is displayed and Max inflate stops and the
system returns to the previous settings . "Start" must
be pressed to resume therapy .
WARNING
Deflate the XPRT™ therapeutic support surface or use Max Inflate to inflate it completely before beginning CPR .
www.stryker.com
Operation Guide
2950-109-005 REV B
Return To Table of Contents
1-23
English