Emc Information; Xprt™ Therapeutic Support Surface - Stryker XPRT Manual De Uso

Superficie de apoyo terapéutico. integración con fl27 (2155/2156)
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 293
English
XPRT™ THERAPEUTIC SUPPORT SURFACE
The XPRT™ is suitable for use in the electromagnetic environment specified below . The customer or the user of the
XPRT™ should assure that it is used in such an environment .
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrostatic fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips,
variations
and
interruptions on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a .c . mains voltage prior to applications of the test level .
T
Return To Table of Contents
1-68

EMC Information

Guidance and Manufacturer's declaration - Electromagnetic Immunity
IEC 60601 Test Level
+6 kV contact
+8 kV air
+2 kV for power
supply lines
+1 kV for input/
output lines
+8 kV differential mode
+2 kV common mode
voltage
<5% Ut (>95% dip in Ut)
short
for 0 .5 cycle
40% Ut (60% dip in Ut) for
5 cycles
70% Ut (30% dip in Ut) for
25 cycles .
<5% Ut (>95% dip in Ut)
for 5 seconds
3 A/m
Compliance Level
+6 kV contact
+8 kV air
+2 kV for power
supply lines
+1 kV for input/
output lines
+8 kV differential mode
+2 kV common mode
<5% Ut (>95% dip in Ut)
for 0 .5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles .
<5% Ut (>95% dip in Ut)
for 5 seconds
3 A/m
2950-109-005 REV B
Electromagnetic
Environment Guidance
Floors
should
be
wood,
concrete,
or
ceramic
tile .
If floors are covered with
synthetic material, the relative
humidity should be at least
30% .
Main power quality should be
that of a typical commercial or
hospital environment .
Main power quality is that of
a typical commercial and/or
hospital environment .
Main power quality should be
that of a typical commercial
and/or hospital environment .
XPRT™
If the user of the
requires continued operation
during power main interruptions,
it is recommended that the
device be powered from an
uninterrupted power supply or
a battery .
Power
frequency
magnetic
fields should be at levels
characteristic
of
a
typical
location in a typical commercial
and/or hospital environment .
www.stryker.com

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2950

Tabla de contenido