To Access The Service Screen - Stryker XPRT Manual De Uso

Superficie de apoyo terapéutico. integración con fl27 (2155/2156)
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 293
English
WARNING
• Explosion risk . Do not use in the presence of flammable mixture with air or with oxygen or nitrous oxide .
• Electrical shock risk . Refer all servicing to qualified personnel .
• The service screen and its functions are for use by authorized service personnel only otherwise equipment
malfunction could result .

TO ACCESS THE SERVICE SCREEN

1 .
Connect a pendant controller to the black connector
on the foot box .
Note: XPRT support surfaces integrated with InTouch 1 .x
and 3 .x do not require the pendant to access the service
screen . However, XPRT with 2 .x versions of the software
require the pendant controller to access the service
information .
2 .
Plug the support surface power cord into a power
source .
3 .
Push the Support Surfaces button/tab .
4 .
With the support surface and touch screen
functioning:
a) Select the lock button
b) Select and hold the "Percussion" button
for approximately five seconds .
c) Select and hold the "Rotation" button for
approximately five seconds until the service
welcome displays on the footboard display .
5 .
Select the "Start" button to enter the service screen .
Return To Table of Contents
1-54
Service Information
.
2950-109-005 REV B
www.stryker.com

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2950