English
TOP COVER REPLACEMENT
1 .
Unplug the support surface power cord from the power source .
2 .
Unzip the support surface cover at the patient's right side midsection and remove .
3 .
Line up the new cover zipper with the bottom cover zipper and zip them together .
4 .
Plug the support surface power cord into the power source and test the functionality before returning to service .
FOOT BOX POWER BOARD REPLACEMENT
Required Tools:
• Phillips Screwdriver
• ESD System (Static Strap)
1 .
Unplug the support surface power cord from the power source .
2 .
Fold the foot section of the support surface back over the head section .
3 .
Using a Phillips screwdriver, remove the seven Phillips head screws holding the cover to the foot box assembly
and remove the cover .
4 .
Properly ground yourself .
5 .
Unplug the cable connections from the power board . Note the locations so the new power board will be connected
properly .
6 .
Using a Phillips screwdriver, remove the six screws holding the power board to the foot box assembly .
7 .
Remove the power board and contact your local Stryker service representative to return for fault evaluation .
8 .
Reverse the procedure to install the new power board .
9 .
Plug the support surface power cord into the power source and test the functionality before returning to service .
FOOT BOX TRANSFORMER REPLACEMENT
Required Tools:
• Phillips Screwdriver
• ESD System (Static Strap)
• Small Regular Screwdriver
1 .
Unplug the support surface power cord from the power source .
2 .
Fold the foot section of the support surface back over the head section .
3 .
Using a Phillips screwdriver, remove the seven Phillips head screws holding the cover to the foot box assembly
and remove the cover .
4 .
Properly ground yourself .
5 .
Using a small regular screwdriver, unplug the cable connections from the junction block . Note the locations so the
new transformer will be connected properly .
6 .
Using a Phillips screwdriver, remove the screw from the bottom of the foot box securing the transformer to the foot
box assembly .
7 .
Remove the transformer and contact your local Stryker service representative to return for fault evaluation .
8 .
Reverse the procedure to install the new transformer .
9 .
Plug the support surface power cord into the power source and test the functionality before returning to service .
Return To Table of Contents