13. Per hospital protocol, record the patient's vital signs and ensure adequate distal perfusion is maintained. If necessary,
adjust air volume in balloon.
NOTE: Air volume and compression time may differ according to patient's condition, anticoagulant dosage, and size of
puncture site.
Device Deflation and Removal
1. If desired, all air may be removed slowly at the appropriate amount of time to achieve complete hemostasis; or, a small
amount of air (such as 1-2 mL) may be removed from the balloon periodically based on the hospital's preferred clinical
technique/hospital protocol to aid in maintaining hemostasis.
NOTE: If PreludeSYNC DISTAL syringe is not available during air removal or re-injection, the cap on the valve may be
removed by twisting and a standard luer syringe can be attached.
CAUTION: DO NOT remove cap while sheath is still in patient.
2. Upon reaching complete hemostasis, slowly withdraw remaining air from the PreludeSYNC DISTAL and observe the
access site for bleeding.
NOTE: Complete hemostasis time may differ according to patient's condition, anticoagulant dosage, and size of
puncture site.
3. If bleeding is present, inject air until bleeding stops. Wait sufficient time and repeat the device deflation and removal
instructions.
4. Once the air is removed and hemostasis is confirmed, carefully remove the PreludeSYNC DISTAL.
5. Place a sterile dressing per hospital protocol.
Caution: Federal (USA) law restricts this
device to sale by or on the order of a
physician.
Single use.
Do not use if package is damaged.
Left hand device
L
Medical Device
For electronic copy scan QR code or go to www.merit.com/ifu and enter IFU ID Number. For printed copy,
call U.S.A or E.U Customer Service.
Caution: Consult accompanying documents.
Read instructions prior to use.
Do not re-sterilize
Sterilized using ethylene oxide.
Right hand device
R
Sterile
Sterile Package
Package