5.3 Connecting the foot switch (optional)
Only the foot switch cited in chapter "4.3 Accessories and
spare parts" on page 37 will ensure safe operation.
At first delivery, the "CONTINUAL" footswitch mode is se-
lected (see chapter "6.2.2 Selecting the foot switch mode"
on page 43).
Insert the jack plug for the foot switch into the socket for
E
the
foot switch (
) on the back of the control unit.
Fig. 4: Socket for the foot switch
5.4 Connecting the handpiece
If no handpiece is connected to the controller, a message
appears on the display:
Handpiece
detached!
Plug in to start
Fig. 5: Handpiece detached
Connect the handpiece cable jack socket supplied in the
E
scope of delivery to the connection on the handpiece of
the device.
Insert the FireWire plug of the handpiece cable fully into
E
the output socket of the controller.
Fig. 6: Output socket
The handpiece is detected by the controller and the pre-
set puncturing frequency appears on the display.
40
EN
If the handpiece is not detected:
Refer to chapter "10 Questions and problems" on
E
page 47.
5.5 Cleaning and disinfecting the
equipment
ATTENTION
Damage due to fluids
If cleaning products or disinfectant gets inside the hand-
piece, the controller or the power supply, this may cause
a short-circuit. In addition, this may also corrode the elec-
trical and mechanical components.
Never immerse the handpiece, the controller or the
E
power supply in cleaning product or disinfectant.
Never clean the handpiece, the controller or the
E
power supply in autoclaves or in an ultrasonic clean-
ing bath.
Consider material compatibilities when selecting
E
cleaning materials and disinfectants (see chapter
"7.2 Material compatibility" on page 45).
ATTENTION
Device damage caused by unapproved cleaning
materials or disinfectants
Cleaning materials or disinfectants that are not compat-
ible with the materials of the device can cause damage to
its surface.
Use only cleaning materials and disinfectants that are
E
authorised for use by the legislation in your country.
Consider material compatibilities when selecting
E
cleaning materials and disinfectants (see chapter
"7.2 Material compatibility" on page 45).
Disconnect the controller from the power supply.
E
Wipe the controller, the handpiece cable, the handpiece
E
and the handpiece tray with a soft cloth moistened with
cleaning product.
Wipe the handpiece, the controller, the handpiece cable
E
and the handpiece tray with a soft cloth moistened with
disinfectant.
EXCEED