FAULT/
FAULT
MESSAGE
1 1 M A N U F A C T U R E R D E C L A R A T I O N S
11.1 Guarantee
This device is a high-quality medical device.
The product is covered by the statutory 2-year warranty
against malfunctions due to material failure or production
defects.
We guarantee that the safety cartridges will remain sterile
for 5 years in sealed and undamaged packaging, if the
transport and storage conditions are complied with. In the
event of complaints relating to safety cartridges, please
provide us with the lot number printed on the label.
The warranty does not apply to the following damage:
Damage and consequential damage caused by improper
O
use or failure to comply with the operating manual.
Damage caused by the ingress of fluids or dirt into the
O
inside of the handpiece, the controller or the power sup-
ply.
11.2 Electromagnetic compatibility
WARNING
The device has been tested for electromagnetic compatibil-
ity (EMC) and is suitable for use in the area of domestic
healthcare (CISPR 11 class B). Use in motor vehicles and
aeroplanes is not permitted.
Caution when using this device next to or in connection with
other devices. This could lead to defective operation.
48
EN
RECTIFICA-
CAUSE
TION
or standby
mode" on
page 42).
Switch the
E
controller
back to oper-
ating mode.
If it is not
E
possible to
establish
standby
mode, discon-
nect the
power supply
FAULT/
FAULT
MESSAGE
In order to rectify further functional disruptions:
Firstly, disconnect all components, such as the hand-
E
piece and the power supply, from the controller.
Check all connections and reconnect the components.
E
Check the function of the device again.
E
If malfunctions persist or if you have any questions or
E
complaints, please contact your specialist retailer.
For our current offers, a selection of safety cartridges and
accessories, visit www.amieamed.com.
Visit www.amieamed.com/contact-us/distributors to find
your local dealer and related contact information.
If such use or installation is necessary, this device and the
other devices must be checked to ensure they are function-
ing normally.
When using accessory parts, power units and cables that
are not included in the "4.3 Accessories and spare parts" on
page 37 of the device manufacturer, there is a risk of in-
creased electromagnetic emissions or reduced electromag-
netic immunity of this device and therefore defective opera-
tion.
Portable HF communication devices (such as telephones,
tablets, etc.) should not be used at distances of less than
30 cm from the device and its electrical accessories. Other-
wise this could impair/adversely affect the capacity of the
device.
With regards to guaranteeing its basic safety and significant
performance characteristics, the device has been tested for
its electromagnetic immunity and its electromagnetic radi-
ation. These remain within the limit values specified in the
following tables, with consideration to the stipulated min-
imum distance from neighbouring devices. No maintenance
measures are required to guarantee basic safety and the
significant performance characteristics.
11.2.1 Manufacturer declarations on
electromagnetic emissions
The device is intended for operation in the electromagnetic
environment specified below. The user of the device should
ensure that it is operated in this type of environment.
RECTIFICA-
CAUSE
TION
from the con-
troller for 30
seconds.
EXCEED