Precision™ System Clinician Manual
Note: If you accidentally locate the patch
in the wrong place, or if the Charging
Belt moves out of alignment, the
Charger will start beeping again. Use
a new adhesive patch or readjust the
belt to place the Charger back into
position.
5. When the Charger emits a series of
double beeps, the Stimulator is fully
charged. Turn off the Charger, remove
the Charging Belt or adhesive patch, and
return the Charger to the Base Station.
Do not confuse the end of charge signal
(a series of double beeps) with the
continuous beeps that indicate that the
Charger is searching for the Stimulator.
Note: • Depending on your program
parameters, you may expect daily
recharging times from as low as
10 minutes up to four hours, or
weekly recharging times from as
low as one hour up to four hours.
Charging times may be longer than
4 hours when not properly aligned.
• The end of a charge signal is
a distinct double beep, and the
alignment indicator is a steady
continuous signal.
• Remote will not be able to
communicate with the IPG when
charging.
Clinician Manual
9055959-008 Rev A
40 of 327

Patient Remote Control

Basic Operation
The Remote Control communicates with the
Stimulator through a radio frequency (RF)
telemetry link from a distance of up to 60 cm
(2 feet).
The Remote Control goes into idle (or sleep)
mode when not in use. It can be reactivated
by pressing any button. To unlock the Remote
P
Control, hold down
, until "Release P To
Unlock" appears.