Connecting to the Trial
Stimulator
1. If two leads are used, connect the cable
labeled "1-L" to the upper or left lead,
and the cable labeled "2-R" to the lower
or right lead. Labels are provided. If two
2x4 splitters are used, connect the cable
labeled "1-L" to the proximal connector
of one of the splitters and connect the
cable "2-R" to the proximal connector of
the other splitter. If using a Splitter 2x8
connect the cable labeled "1-L" to the
laser-marked tail and the cable labeled
"2-R" to the unmarked tail.
2. Connect the OR cables to the Trial
Stimulator.
If only one lead is used, connect the OR
cable to "1-L" on the Trial Stimulator.

Connecting to the Trial Stimulator

3. Fit the Trial Belt to the patient, cut off
the excess length, and place the Trial
Stimulator in the belt pocket.
1. OR Cable
2. OR Cable Connector
3. External Trial Stimulator (ETS)
9055959-008 Rev A 31 of 327
4. Trial Belt
5. Lead or Extension