Precision™ System Clinician Manual
For Percutaneous Lead - Single
Lead Connection
• Connect a single lead, extension, or
2x4 splitter to IPG port "1-L".
• Insert port plug into unused port.
For Surgical Paddle Lead
Connection (SCS Only)
• Left side of lead connects to IPG port
"1-L".
• Right side of lead connects to IPG
port "2-R".
Note: The right tail of the Paddle Lead
is laser-etched to allow for easy
identification of right and left.
Left
Right
Upper
1L
2R
1L
1. Fully insert the lead(s), extension(s),
or splitter(s) into the IPG port(s), being
careful not to stress or bend the proximal
end of the lead or extension. When the
lead is properly inserted, the lead will
stop and the retention ring will be located
under the setscrew.
Clinician Manual
9055959-008 Rev A
Lower
Single
2R
1L
34 of 327
Note: If you experience difficulty when
inserting the lead, lead extension,
or port plug, use the hex wrench
to loosen (counterclockwise) the
setscrew and/or gently rotate the lead
to help advance the proximal end.
Note: To confirm good connections, check
impedances before tightening the
setscrew.
2. Pass the hex wrench through the hole in
the septum located on the top of the IPG
header and tighten each set screw until
the hex wrench "clicks," indicating lock.
CAUTION: Ensure that the lead is fully
inserted before tightening the setscrew to
prevent lead damage.