Precision™ System Clinician Manual
P
2. Press the
button to cycle through the
programs and select the program to be
restored.
3. Press p to restore the clinic-programmed
settings.
Coverage Area Selection
Each program can consist of up to 4 different
coverage areas (Areas). The stimulation
strength and other settings for each Area can
be adjusted.
1. Press the
button to access the Area
"
level screen. This controls the stimulation
strength for each Area individually. Press
the
button as needed to highlight a
"
specific stimulation coverage area.
2. Press the p or q button to adjust the
stimulation strength (amplitude) of the
selected Area.
Clinician Manual
9055959-008 Rev A
44 of 327
Additional Area Options
The Pulse Width and Rate parameters are
available as programming options to allow
patients greater control of their therapy. By
default, access to Pulse Width and Rate
are disabled in the patient Remote Control.
Access to these parameters can be enabled
through the Bionic Navigator™ software.
To Access the Pulse Width and Rate Options
(if enabled through Bionic Navigator):
1. Press and hold the
"
Remote Control for three seconds until
the Pulse Width screen appears. This
is indicated by "Width" in the bottom left
corner of the screen.
2. Area 1 will initially be highlighted. Press
the
button to cycle to the Area you wish
"
to adjust.
button on the