Precision™ System Clinician Manual
Note: Avoid common sources of
communication interference such as
televisions and computer monitors.
Note: Communication between the Remote
Control and the Stimulator is not
possible while charging. Temporarily
discontinue charging in order to use
the Remote Control.
Stimulation Amplitude
The Level screen is the Remote Control's
"default" screen.
Press the p or q button from the Level
screen to increase or decrease stimulation
strength.
Clinician Manual
9055959-008 Rev A
42 of 327
Stimulator Battery Status
The Level screen also displays the Stimulator
battery status in the top center of the screen.
Three solid bars represent a fully charged
Stimulator. As the Stimulator battery depletes,
depending on the patient's stimulation
settings and usage, the bars will "empty"
accordingly. For their convenience, patients
are encouraged to recharge the Stimulator at
the first "Recharge Battery Soon" message
displayed by the Remote Control. Due to
Zero Volt™ Technology, the Stimulator can be
completely and repeatedly discharged without
causing battery failure.
Program Selection and Activation
Note: The program that is currently or
was most recently running will be
underlined ( _ ). Empty program slots
are denoted by an empty box [ ] with
no program number.
P
1. Press the
button from the Level screen
to go to the Program selection screen.
Program 1 will be highlighted.