Device Linking - Boston Scientific Precision SC-1110 Serie Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. When the desired Area is highlighted,
press p or q to increase or decrease
the Pulse Width.
4. To continue to Rate adjustment, press
and hold the
button for three seconds
"
from any Pulse Width Area screen. The
Rate screen is indicated by "Rate" in the
bottom left corner of the screen.
5. Area 1 will initially be highlighted. Press
the
button to cycle to the Area you wish
"
to adjust.
6. When the desired area screen is
highlighted, press p or q to increase or
decrease the Rate.
7. To return to the main Level screen, press
and hold the
button for three seconds
"
until the Level screen appears.

Device Linking

A Remote Control can communicate with only
one Stimulator at a time. This prevents the
Remote Control from accidentally controlling
an unintended device. In the linking process,
the Remote Control will identify, by telemetry,
the intended Stimulator for communication.
P
1. Press and hold the
the Remote Control and initiate
communication between the Remote
Control and the Stimulator.
button to unlock
Rechargeable Stimulator System
2. If the Remote Control is not linked to a
Stimulator, it will display a "Link?" screen.
P
Press
to link. The Remote Control will
search for the nearest Stimulator.
If the Remote Control is already linked to
a Stimulator, it may continuously search
for it. This link must be cleared before
a new link can be established. (See
"Clinician Options" on page 47).
3. The Remote Control will identify the
Stimulator by serial number. Verify before
proceeding.
P
4. Press
to confirm and continue. Press
to cancel.
9055959-008 Rev A 45 of 327
"