0210-100-700 Rev-T
Inspection and Testing
WARNINGS:
• DO NOT use any equipment if damage is apparent or the inspection criteria
are not met.
• DO NOT disassemble or service this equipment, unless otherwise specified.
NOTE: For service, contact your Stryker sales representative or call Stryker
customer service. Outside the US, contact your nearest Stryker subsidiary.
INTERVAL
INSPECTION CRITERIA
Before Use
Check equipment for damage,
wear, or missing components.
Check power cord and tubing for
cuts or punctures.
Check the handpiece and battery
power pack housing for cracks.
NOTE: If any component must be discarded, see the Disposal/Recycle section.
Cleaning
(Between Wound Care Use - Single Patient Use Only)
WARNINGS:
• DO NOT immerse any system component in liquid.
• DO NOT sterilize any system component. See the instructions for use supplied
with the InterPulse Power Pack Charger.
CAUTIONS:
• DO NOT use solvents, lubricants, or other chemicals, unless otherwise
specified.
• DO NOT allow liquids or moisture into the electrical connections of the
rechargeable power pack or battery power pack.
1. Remove the tip assembly from the handpiece. Properly discard of the tip
assembly.
2. Wipe the handpiece with a soft cloth dampened with a nonabrasive, hospital
disinfectant.
3. Dry the equipment with a lint-free towel.
4. Install new suction tubing and tip assembly to the handpiece. See the
Instructions section.
Troubleshooting
WARNING: DO NOT disassemble or service this equipment unless
otherwise specified.
NOTE: If you require service, contact your Stryker sales representative or call
Stryker customer service. Outside the US, contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
The handpiece fails to
The handpiece is not
provide suction.
connected to a facility
vacuum source or
the vacuum source is
adjusted incorrectly.
The handpiece is
damaged.
Sporadic electrical
Electrical noise is
interference is
present.
experienced.
NOTE: If any component must be discarded, see the Disposal/Recycle section.
www.stryker.com
Print Date: Nov 02, 2016 02:57:27 PM
0000004059, Rev. T Effective Date: Sep 26, 2016 2:47:57 PM
Storage and Handling
CAUTION: ALWAYS store the equipment within the specified environmental
condition values throughout its useful life. See the Specifications section.
To ensure the longevity, performance and safety of this equipment, use of the
original packaging material is recommended when storing or transporting this
equipment.
Disposal/Recycle
ACTION
If damage is apparent,
• ALWAYS follow the current local recommendations and/or regulations
replace the equipment.
• ALWAYS decontaminate the equipment and battery pack exposed to infectious
To Remove the Battery Pack (figure 3)
• DO NOT remove the battery pack in the presence of a patient.
• DO NOT cut the irrigation tubing/power cord to remove the battery power pack
1. Push the two battery lid removal tabs and remove the lid from the battery
2. Disconnect the leads.
3. Remove the battery pack.
4. Recycle or dispose of the battery pack and equipment properly.
SOLUTION
Make sure the handpiece
is connected to a facility
vacuum source and
the vacuum regulator is
adjusted correctly.
Replace the handpiece.
Turn off all the electrical
equipment not in use in the
room.
Relocate the electrical
equipment to maximize
the distance between the
equipment. Increase spatial
distance.
Plug equipment into
different outlets.
WARNINGS:
governing environmental protection and risks associated with recycling or
disposing of the equipment at the end of its useful life.
material before sending them to a waste treatment facility.
To comply with European Community Waste Electrical and
Electronic Equipment (WEEE) Directive 2012/19/EU, this
device should be collected separately for recycling. Do
not dispose of as unsorted municipal waste. Contact local
distributor for disposal information. Ensure infected equipment
is decontaminated prior to recycling.
To comply with European Community Batteries Directive
2006/66/EC, this device has been designed for safe removal
of the batteries by the end user using the instructions provided
herein. Infected units and batteries should be decontaminated
before they are sent for recycling.
WARNINGS:
from the handpiece. Failure to comply may cause electrical shock, excessive
heat and/or sparks and result in injury and/or fire.
power pack housing.
Figure 3 – To Remove the Battery Pack
EN
5