Specifications (continued)
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment specified below.
The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series should assure that it is used in such an
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency (50/60 Hz) magnetic
field
IEC 61000-4-8
Voltage dips, short interruptions and
voltage variations on power supply input
lines
IEC 61000-4-11
NOTE: U
is the alternating current mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the
The InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series as recommended below, according to the maximum output power of the
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
8
Print Date: Nov 02, 2016 02:57:27 PM
0000004059, Rev. T Effective Date: Sep 26, 2016 2:47:57 PM
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
3 A/m
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
InterPulse Irrigation System, REF 0210-XXX-000E Series and REF 0211-XXX-000E Series
communications equipment.
150 kHz to 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
environment.
Compliance level
±2 kV, ±4 kV, ±6 kV
Floors should be wood, concrete or ceramic tile. If floors
contact
are covered with synthetic material, the relative humidity
±2 kV, ±4 kV, ±8 kV air
±2 kV on DC power
lines
Not Applicable
characteristic of a typical location in a typical commercial
3 A/m
Not Applicable
Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
Electromagnetic environment - guidance
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Not Applicable
Power frequency magnetic fields should be at levels
or hospital environment.
Not Applicable
800 MHz to 2.5 GHz
d=2.3√P
0210-100-700 Rev-T
0.23
0.73
2.3
7.3
23