Preparation for Cleaning
1. Remove any light post adapters from the scope prior to
cleaning.
2. Prepare an enzymatic detergent
manufacturer's recommendations.
3. Wipe the entire device with the detergent, using a clean
cloth.
4. Immerse the device in the detergent. Inject any inside
regions of the device with 50 mL of the detergent solution
to ensure all parts of the device are reached.
5. Soak the device in the detergent for 15 minutes.
Manual Cleaning
1. Brush
• Prepare a fresh solution of enzymatic detergent
according to the manufacturer's recommendations.
• Thoroughly brush the exterior of the device with a soft-
bristled brush, focusing on any mated or rough surfaces.
• Using a syringe, inject any mated surface 5 times with
50 mL of the detergent.
2. Rinse
• Rinse the device with water
until all detergent residue is removed.
• Flush any mated surfaces 5 times.
• Once all detergent residues is removed, continue to rinse
for a minimum of 30 seconds.
• Drain excess water from the device and dry it using a
clean cloth or pressurized air.
• Visually inspect the device for cleanliness, paying close
attention to hard-to-reach areas. If visible soil remains,
repeat steps 1 and 2.
3. Soak
• Prepare a non-enzymatic detergent
manufacturer's recommendations.
• Fully immerse the device and inject any mated surfaces
with 50 mL of the detergent.
• Soak the device for 15 minutes.
4. Brush
• Thoroughly brush the exterior of the device using a
soft-bristled brush.
• Inject the prepared detergent into any mated surfaces
5 times.
5. Rinse
• Thoroughly rinse the device with water
detergent residue is removed. Flush any crevices 5 times.
Once all detergent residue is removed, continue to rinse
for 30 seconds.
• Drain the excess water from the device.
1
Enzol® at 1 oz/gal (US) at 35 – 40°C was validated for cleaning efficacy.
2
Reverse Osmosis/Deionized (RO/DI) water was validated for cleaning efficacy.
3
Prolystica 2x Neutral Detergent at 1/8 oz/gal was validated for cleaning efficacy.
Automated Cleaning
1. Brush
• Thoroughly brush the device with an appropriately sized
soft-bristled brush, focusing on hard-to-clean areas.
2. Rinse
• Rinse the device with water
until all visible detergent residues have been removed.
• Flush all crevices and mated surfaces at least five times.
• Continue to rinse the device for a minimum of
30 seconds after all detergent residue has been removed.
EN-4
according to the
1
at ambient temperature
2
, according to the
3
until all
2
at ambient temperature
2
1