5.6 THE HANDLE RELEASE CLIP IS NOT ATTACHED TO THE DELIVERY SYSTEM HANDLE

Potential Causes
A. The red release clip has become loose, but is
still attached to the handle, either during transit
or use.
B. The red release clip is no longer attached to
the release wire.
Recommended Actions
A.1. Check that the red release clip is still
attached to its release wire by gently pulling
it to ensure that tension can still be felt.
A.2. If still attached, the red release clip can be
pushed back into its housing in the handle.
A.3. The red release clip can be removed and
retracted as per the IFU section 3.5.4.
B.1. Remove the release wire
B.2. The release wire should be retracted as
described in the IFU section 3.5.4 using
atraumatic artery forceps. The atraumatic
artery forceps should be attached to the
release wire and retracted.
Note: Only one wire is present in the
delivery system
21