Aesculap
High-frequency surgical device GN300
Error
Error cat-
Error
code
egory
description
Er 28
Applica-
Foot switch
tion/
defective
Monopolar
activation
Er 30
Applica-
Time limit
tion/
exceeded
Bipolar
activation
Er 31
Applica-
Temperature
tion/
monitoring
Bipolar
of bipolar
activation
final stage
Er 32
Applica-
Temperature
tion/
monitoring
Bipolar
of bipolar
activation
final stage
Er 33
Applica-
Dosage
tion/
monitoring
Bipolar
activation
Er 38
Applica-
Foot switch
tion/
signals
Bipolar
defective
activation
Er 41
Internal
HF switch-
check!
on monitor-
Hardware
ing
Er 42
Internal
HF switch-
check!
on monitor-
Hardware
ing
Er 400
Internal
Hardware
check!
error
Hardware
Er 500
Internal
Software
check!
error
Software
ER600
Internal
Incompati-
check! Cal-
ble software
ibration
adjustment
24
®
Message
Effects on the HA output
displayed
monopolar
bipolar blocked
monopolar possible
bipolar
bipolar blocked
monopolar possible
bipolar
bipolar blocked
monopolar possible
bipolar
bipolar blocked
monopolar possible
bipolar
bipolar blocked
monopolar possible
bipolar
bipolar blocked
monopolar possible
monopolar
monopolar blocked
bipolar possible
bipolar
bipolar blocked
monopolar possible
monopolar
monopolar blocked
bipolar blocked
monopolar
monopolar blocked
bipolar blocked
monopolar
monopolar possible
bipolar possible
Suggested remedy
Comments
Inspect and replace foot
To start HF mode, use handpiece
switch
with finger switches if neces-
sary
Release the pedal
HF can be activated for max.
30 seconds for each activation
of the pedal
Check cable and instru-
Only use approved accessories!
ment
Allow device to cool
The thermal time constant
down
means that the device can only
be reactivated after approx.
2 minutes
Switch off the unit and
If the error message is displayed
switch it on again
again after switching the device
on and off, the device must be
handed over for repair
Check the foot switch
If an error is still displayed, the
foot switch is defective
Switch off the unit and
Error is displayed for non-acti-
switch it on again
vated monopolar HF
Switch off the unit and
Error is displayed for non-acti-
switch it on again
vated monopolar HF
Switch off the unit and
If the error message is displayed
switch it on again
again after switching the device
on and off, the device must be
handed over for repair
Switch off the unit and
If the error message is displayed
switch it on again
again after switching the device
on and off, the device must be
handed over for repair
Have device checked as
Continued use possible
soon as possible
Limitations in the area of out-
put value accuracy