Models MCA66S-3 | MCA66P-3
Scale interval (d)
Maximum capacity (Max)
Repeatability at 5% load
Standard deviation of the load values, tolerance
Repeatability at approx. maximum capacity
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
Linearity deviation
Tolerance
Typical value
Deviation when load is off-center, positions according to OImL R76
Test weight
Tolerance
Typical value
Sensitivity drift of +10°C to +30°C
Tare-maximum capacity: Less than 100% of maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive 2014/31/EU
Minimum load (Min) according to Directive 2014/31/EU
Minimum initial weight according to USP (United States
Pharmacopeia), Chap. 41
Optimum minimum initial weight
Typical stabilization time
Typical measurement time
MCA66S-3
Unit
Value
mg
0.001
g
61
mg
0.0015
mg
0.004
mg
0.0025
mg
0.02
mg
0.005
g
20
mg
0.02
mg
0.01
ppm/K
1
I
mg
1
mg
0.1
mg
0.82
s
3.5
s
10
Operating Instructions Cubis
Technical Data
MCA66P-3
Value
0.001 | 0.01
12 | 61
0.002
0.01
0.006
0.02
0.008
20
0.03
0.012
1
I
1
0.1
0.82
3.5 | 2.5
10 | 6