Operation
60
Operating Instructions Cubis
8.3.2
Opening or Closing via Proximity Sensors
The proximity sensor works in "crossover mode":
— Left proximity sensor: Opens and closes the right-hand and the upper
door
— Right proximity sensor: Opens and closes the left-hand and the upper
door
The sensitivity of the proximity sensors can be adjusted (see
Chapter "4.13.2 "Settings" Menu", page 36). Wearing safety gloves may
inhibit the use of the proximity sensor.
Requirements
A license key has been entered to use the motorized draft shield.
Procedure
Hold your hand in front of the left or right proximity sensor. This enables
t
the door to be fully opened or closed.
8.3.3
Opening or Closing a Motorized Draft Shield at the
Door Handle
Procedure
To open or close a door: Press on the door handle (1) of a door. The door
t
is automatically fully opened or closed.
1
8.4
Leveling the Device
Leveling compensates any inclines at the device's installation site. If leveling
is necessary: The [Leveling] button appears in the weighing display and a
message appears in the status center.
Procedure
If the weighing display is displayed: Press the [Leveling] button.
t
When the status center is displayed: Press the [Level] button.
t
The Leveling Wizard opens.
y
Follow the wizard's instructions.