Procedure
Open the "Settings" / "Device Settings" / "Device Information" menu.
t
Select the "View data storage device" menu item.
t
A list of all the values saved in the data storage device is displayed.
y
If only the values for a desired date need to be displayed:
t
Press the [Filter] button.
t
Type the desired date into the entry field.
t
To filter the list with the date entered: Press the [OK] button.
t
If the list needs to be searched for a particular ID:
t
Press the [Search] button.
t
Type the desired ID into the entry field.
t
To start the search with the entered ID: Press the [OK] button.
t
8.11
Starting the Ionization Process
Requirements
— The application for use of the ionizer is activated.
— The [Ionizer] button appears in the operating display.
Procedure
Press the [Ionizer] button.
t
8.12 Running Applications (Examples)
8.12.1
Executing the "Unit change" Function
The "Unit change" function enables the user to switch between the
different units and resolutions defined in the weighing profile of the active
task. The units and resolutions can be set at the beginning of the weighing
process.
Procedure
Start the desired task.
t
Press the [Unit change] button.
t
All units defined in the weighing profile for the active task are displayed
y
in a list.
All resolutions for the weight value defined in the weighing profile for the
y
active task are displayed in a list.
Press the desired unit.
t
To set the resolution for the selected unit: Press the desired resolution.
t
To confirm the selection and return to the weighing display: Press the
t
[OK] button.
The current weight value is displayed in the selected unit and resolution.
y
Operating Instructions Cubis
Operation
| MCA Models
67
®