Troubleshooting

Problem
"Restart Camera
Console" message
(Color bar
background)
"System Error"
message (Light blue
background)
No color bar
No color bar
(Optical DVI only)
Incorrect picture
color
White balance
(WB) quality not
good
Picture is too dark
Picture is too bright •
Noise or snow on
picture when using
electrocautery
probes
Noise or snow
on picture
when not using
electrocautery
probes
28
Possible Solution
Camera head temporarily shut down due to overcurrent. Turn off the
console, wait 3 seconds, and turn it back on.
After sterilization, ensure the camera head has cooled down before
connecting it to the console
No video detected.
After sterilization, ensure the camera head has cooled down before
connecting it to the console
Return the system for repair.
Ensure the video-out from the console is connected to the video-in on
the monitor.
Ensure all video systems are powered on.
Ensure that the camera head is not connected to the console.
Turn off the console, wait 3 seconds, and turn it back on.
Same as above.
See the "Using the DVI Fiber Outputs" section of this manual.
Perform the white balance procedure. (See the "W Button" section of
this manual.)
Check the color settings on the monitor.
See the solution for "Picture is too dark. "
See the solution for "Picture is too bright. "
Perform the white-balance procedure with the light source connected
to the scope. Use metal-halide, xenon, or LED lighting (no fluorescent
lighting).
Increase the camera light level with the camera head.
Increase the light-source output.
Check the fiber-optic light cable for excessive broken fibers.
Decrease the camera light level.
Decrease the light-source output.
Ensure that Shutter submenu in the Configuration menu has the
following settings:
BRT CNTRL AUTO
Shutter On
Plug the electrocautery generator into a separate electrical outlet and
separate the 1288 HD power cord from the electrocautery power cord.
Separate the camera cable from the electrocautery cable.
Reposition the electrocautery grounding pad on the patient.
Reduce Enhancement.
Check for and replace faulty video cables.